Service Cases

Park Atomic Force Microscope NX20

2025/12/05

Customer Information: Wuhan University of Technology

Equipment Information: PARK Atomic Force Microscope NX20

Equipment Failure: AFM controller motherboard failure, indicator light off and no PSPD signal

Fault Diagnosis: Confirmed that the fault is caused by a component failure related to the 15V power supply

Repair Plan: Replace the faulty components and clean the board; test